Development of PSA Process for Medical Oxygen Generator 


Vol. 15,  No. 2, pp. 75-80, Jun.  2009


PDF
  Abstract

In order to separate oxygen from air, the effects of feed gas flow rate and rinse gas flow rate on the product purity and flow were examined using 2 bed PSA with 4 step cycle. The addition of product pressurization step increased the product purity and flow rate. The addition of pressure equalization increased the product flow rate. The test product was manufactured and the purity and flow rate of product oxygen was examined. The results were compared with the commercial medical oxygen generator of 5 ℓ/min and 90% oxygen purity.

  Statistics
Cumulative Counts from November, 2022
Multiple requests among the same browser session are counted as one view. If you mouse over a chart, the values of data points will be shown.


  Cite this article

[IEEE Style]

C. JW and N. BK, "Development of PSA Process for Medical Oxygen Generator," Clean Technology, vol. 15, no. 2, pp. 75-80, 2009. DOI: .

[ACM Style]

Choi JW and Nat BK. 2009. Development of PSA Process for Medical Oxygen Generator. Clean Technology, 15, 2, (2009), 75-80. DOI: .